Bio-Arm

Model: RFM-800

Maternal/Fetal
Product Name:Maternal/Fetal Monitor
Model:RFM-800
P/N:500110
Product's Video:No Video
Inquiry:


Features:            

  • 8.4 inch super resolution color TFT display
  • Maximum 9 waveforms
  • Use built-in system, high integrated intelligent module technology, full digital monitoring, super functions, lowest * swearing and low failure rate
  • Combined functions of maternal monitor and fetal monitor
  • Measure the intrapartum patient and fetal heart rate
  • Uterine contraction and fetal movement marker
  • Continuous real time recording of fetal heart rate
  • Can be switched to normal patient monitoring interface
  • 72 hours graphic trends and tabular of all parameters
  • 1440 groups of NIBP data storage
  • ST analysis
  • Anti-high frequency electronic surgical unit and anti-defibrillation, for monitoring in the surgery or for cardiac care
  • Use wide-angle, multi-wafer, wide beam and ultrasonic probes

Standard configuration:            

  • FHR, Uterine Contraction Pressure(TOCO), FM and Maternal ECG, NIBP, SPO2, PULSE, Built-in rechargeable battery, thermal printer

Optional:            

  • Dual TEMP

Technical Specifications:

Safety

IEC60601 approved (GB9706)

Power source

AC100V-240V, 50/60Hz

Size

330mm(L) × 275mm(W) × 160mm(H)

Net weight

5.0 kg

Working

Temperature: 5 ~ 40
Humidity <= 85%

Transportation and Storage

Temperature: -20 ~ 55
Humidity <= 85%

Ultrasonic Probe: (Fetal Heart Rate)

Type

Type B proof, dirctional, using pulse doppler system

Working frequency

2.5 MHz

Transmitting power density

1.2m W/cm2

Probe

Multi-wafer probe

Sample rate

5 ms

Calculating precision of FHR

0.25 bpm

Accuracy of FHR

1 bpm (100-180 bpm)

Marker (Fetal Movement)

Manually push button mode

When pressed, a red strip will be displayed above the FHR graph

Uterine Contraction Pressure

Type B proof

Uterine contraction pressure probe

Manual zero adjustment

Design pressure sensitivity: 0-100%

ECG

Lead

5 leads (R, L, F, N, C)

Lead selection

I, II, III, avR, avL, avF, V

Waveform

7 channels

Gain selection

×0. 5, ×1, ×2

HR and alarm range

Adult: 15-300bpm
Neonate/pediatrics: 15-350bpm

Accuracy

±2% or ±2bpm, whichever is greater

Input impedance

> 5ΜΩ

Voltage range of electrode polarization

±300mV

Leakage current

< 30uA

Filter

Diagnostic mode: 0.05 ~ 100Hz
Monitoring mode: 0.5 ~ 35Hz
Surgical mode: 1 ~ 25Hz

Scaling signal

1mV ± 3%

Measurement range

-1.0mV ~ +1.0mV

Alarm Range

-0.80mV~0.80mV

RESP

Model

RA-LL impedance

Measurement range

Adult: 0 ~ 120bpm
Neonate/pediatrics: 1 ~ 150bpm

Resolution

1rpm

Accuracy

±2rpm

NIBP

Method

Automatic oscillometric

Operation mode

Auto/Manual

Alarm Type

Systolic, Diastolic, Mean

Measurement ranges

Systolic BP: 40 ~ 300mmHg
Diastolic BP: 10 ~ 220mmHg
Mean BP: 20 ~ 240mmHg

Resolution

1mmHg

Accuracy

Max average error ±5mmHg or ±5%, whichever is greater

Overload Protection

Double safety protection

 SPO2

Measurement Range

0 ~ 100%

Alarm range

0 ~ 100%

Resolution

1%

Accuracy

±2% (80~100%)
±3% (60~79%)

 PULSE RATE

Range

0 ~ 254bpm

Resolution

1bpm

Accuracy

±1bpm or ±2%, whichever is greater

Alarm delay time

10sec

 TEMP

Channel

2 channels

Measurement range

0~50

Alarm range

0~45

Resolution

0.1

Accuracy

±0.2